SPECIFICATIONS
Energy Measure:
Load: 50 Ohms ± 1%
Range: Low 0 - 100 Joules High 0 - 700
Joules
Resolution: Low 0.01 Joules High 1
Joules
Accuracy: 2% of Reading plus ±1
Joules < 100 ±2 Joules > 100
Pulse Width: 0 - 65 milliseconds
Cardioversion: Timing window: -250 to
+250 milliseconds.
Waveform:
Normal Sinus @ 60 BPM 1mV Lead I
Real Time: Oscilloscope
Output 1000:1
Mini Phono Jack
Expanded Time:
Defibrillator waveform expanded 200:1 and is repeated every 8 seconds. The waveform can be
viewed via high level output, ECG leads and paddles.
ECG:
ECG waveform @ 60 BPM is internally generated through the paddles or ECG Output Leads when
Cardio/ Energy selector is set to Cardio.
Power Requirements: 9 Volt Alkaline
Battery or AC Adapter
Dimensions: 9.5"L x 6.25"W x
1.5"H
Case:
Rugged ABS Plastic Case with Stainless Steel internal Shielding Weight: 2lbs. With