Pediatric sizes available

click to view
|
|
|
|
Sensor Technology |
Piezo crystal film |
| Power Requirements |
No external power is required,
signal is self generating |
| Typical Polygraph Settings |
Sensitivity: 2 - 30 uV/mm Low
frequency filter/Time constant: .15 Hz or lower/1 second or longer High frequency filter:
15 Hz |
| Typical Signal Amplitude |
600 microvolts peak to peak.
Typical signal amplitudes may vary considerably. Response is dependent on such variables
as sensor placement, patient airflow, sleep position, polygraph characteristics, etc. |
| Output Impedance |
Static (near DC): 750 KOhms
Dynamic (.25 Hz): 600 Kohms Non-referenced, bipolar output |
| Standard Connector Type |
1.5 mm touch-proof safety
connector |
| Wire Length |
7.0 feet |
| Note: Crystal
Trace® effort sensors are non-referenced and produce bipolar microvolt level signals.
Proper signal processing requires the use of AC coupled, referenced differential
amplifiers |